What is pulse oximetry?
Pulse oximetry is a simple, relatively cheap and non-invasive technique to monitor oxygenation. It monitors the percentage of haemoglobin that is oxygen-saturated.
Oxygen saturation should always be above 95%, although in those with long-standing respiratory disease or cyanotic congenital heart disease, it may be lower, corresponding to disease severity.
The oxyhaemoglobin dissociation curve becomes sharply steep below about 90%, reflecting the more rapid desaturation that occurs with diminishing oxygen partial pressure (PaO2).1 On most machines the default low oxygen saturation alarm setting is 90%.
Pulse oximetry does nicht provide information on the oxygen content of the blood or on ventilation. Thus care is needed in the presence of anaemia and in patients developing respiratory failure due to carbon dioxide retention, for example.
Principles of pulse oximetry
Zurück zum InhaltOximeters work by the principles of spectrophotometry: the relative absorption of red (absorbed by deoxygenated blood) and infrared (absorbed by oxygenated blood) light of the systolic component of the absorption waveform correlates to arterial blood oxygen saturations.2 Measurements of relative light absorption are made multiple times every second and these are processed by the machine to give a new reading every 0.5-1 second that averages out the readings over the last three seconds.
Two light-emitting diodes, red and infrared, are positioned so that they are opposite their respective detectors through 5-10 mm of tissue. Probes are usually positioned on the fingertip, although earlobes and forehead are sometimes used as alternatives. One study has suggested that the ear lobe is nicht a reliable site to measure oxygen saturations.3 However, a more recent study advocated their use in patients admitted to intensive care units for coronary artery bypass surgery.4 Probes tend to use 'wrap' or 'clip' style sensors.
Lesen Sie unten weiter
Central cyanosis, the traditional clinical sign of hypoxaemia, is an insensitive marker occurring only at 75-80% saturation. Consequently, pulse oximetry has a wide range of applications including:
Individual pulse oximetry readings - can be invaluable in clinical situations where hypoxaemia may be a factor - for example, in a confused elderly person.
Continuous recording - can be used during anaesthesia or sedation, or to assess hypoxaemia during sleep studies to diagnose obstruktive Schlafapnoe. Peri-operative monitoring has not, however, been shown to improve surgical outcomes.5
Pulse oximetry can replace blood gas analysis in many clinical situations unless PaCO2 or acid-base state is needed. It is cheaper, easier to perform, less painful and can be more accurate where the patient is conscious (hyperventilation at the prospect of pain raises PaO2).
Pulse oximetry allows accurate use of O2 and avoids wastage. For example, in patients with respiratory failure, rather than limit the use of O2 to maintain hypoxic ventilatory drive, it can be adjusted to a saturation of ~90% which is clinically acceptable.
Neonatal care - the safety limits for oxygen saturations are higher and narrower (95-97%) compared to those for adults.6. A UK pilot concluded that many babies found to have a low oximetry reading were either normal or had a non-cardiac cause for their low oxygen level.7Recent research supports the use of oximetry to screen for congenital heart disease.8
Intrapartum fetal monitoring - the use of fetal pulse oximetry in combination with routine cardiotocography (CTG) monitoring has been studied and found not to reduce the operative delivery rate.9 Future work will look for associations between fetal pulse oximetry use and adverse perinatal and long-term developmental outcomes.10
Pulse oximeters are now used routinely in critical care, anaesthesiology, and A&E departments, and are often found in ambulances. They are an increasingly common part of a GP's kit. Pulse oximetry's role in primary care may include:
Diagnosing and managing a severe exacerbation of chronisch obstruktive Lungenerkrankung (COPD) in the community.
Grading the severity of an Asthma attack. Where oxygen saturations are less than 92% in air, consider the attack potentially life-threatening.11
Assessing severity and oxygen requirements for patients with community-acquired Lungenentzündung.1213
Assessing severity and determining management in infants with Bronchiolitis.
General pointers to the management of hypoxaemia
Oxyhaemoglobin saturation | Behandlung |
90-95% | Measure regularly and especially at night. Review trends. Where value is unexpected, check signal quality and probe. |
80-90% | As above, continuous monitoring and give oxygen until saturations above 90%. |
<80% | As above and consider ventilatory support. |
Using an oximeter2
Zurück zum InhaltResting readings should be taken for at least five minutes.
Poor perfusion (due to cold or hypotension) is the main cause of an inadequate pulse wave. A sharp waveform with a dicrotic notch indicates good perfusion whilst a sine wave-like waveform suggests poor perfusion.
If a finger probe is used, the hand should be rested on the chest at the level of the heart rather than the affixed digit held in the air (as patients commonly do) in order to minimise motion artefact.
Checking that the displayed heart rate correlates to a manually checked heart rate (within 5 beats per minute) generally rules out significant motion artefact.
Emitters and detectors must oppose one another and light should not reach the detector except through the tissue. Ensure the digit is inserted fully into the probe and that flexible probes are attached correctly. Appropriately sized probes should be used for children and infants.
Oximeter accuracy should be checked by obtaining at least one simultaneous blood gas, although this rarely happens. Oximeters may correct average oximeter bias based on pooled data but this does not eliminate the possibility of larger individual biases.
Lesen Sie unten weiter
Sources of error
Zurück zum InhaltPulse oximeters may overestimate true arterial oxygen saturation (SaO2) in people with darker skin tones, particularly black African. The clinical relevance is unclear, but it likely to be greater when SaO2 is lower. 14
Pulse oximetry cannot differentiate between different forms of haemoglobin. Carboxyhaemoglobin is registered as 90% oxygenated haemoglobin and 10% desaturated haemoglobin, thereby causing an overestimation of true saturation levels.2
Significant venous pulsation such as occurs in tricuspid incompetence and venous congestion.
Environmental interference: vibration at 0.5-3.5 Hz and excessive movement. Ambient light - including infrared heat lamps - makes a difference of less than 5%.15
Cold hands - warm extremity if local poor perfusion.
Nail polish should be removed, as it may cause false readings.15
Intravascular dyes, such as methylthioninium chloride, may also temporarily falsely reduce saturation readings.
Improving an oximeter signal16
Warm and rub skin.
Apply a topical vasodilator - eg, glyceryl trinitrate (GTN) cream.
Try an alternative probe site.
Try a different probe.
Try a different machine.
Exclusive updates for healthcare professionals
Stay informed with the latest clinical updates, professional insights, and evidence-based guidance. The Patient Pro newsletter curates essential content for healthcare professionals—delivered straight to your inbox.
By subscribing you accept our Datenschutzrichtlinie. Sie können sich jederzeit abmelden. Wir verkaufen Ihre Daten niemals.
Weiterführende Literatur und Referenzen
- Bhattacharya K; Takuo Aoyagi-a Tribute to the Brain Behind Pulse Oximetry. Indian J Surg. 2020 May 20:1-2. doi: 10.1007/s12262-020-02365-x.
- Gunstone C; Pulse oximetry in primary care. Br J Gen Pract. 2011 Aug;61(589):497. doi: 10.3399/bjgp11X588394.
- Hafen BB, Sharma S; Oxygen Saturation.
- Torp KD, Modi P, Pollard EJ, et al; Pulse Oximetry.
- Haynes JM; The ear as an alternative site for a pulse oximeter finger clip sensor. Respir Care. 2007 Jun;52(6):727-9.
- Seifi S, Khatony A, Moradi G, et al; Accuracy of pulse oximetry in detection of oxygen saturation in patients admitted to the intensive care unit of heart surgery: comparison of finger, toe, forehead and earlobe probes. BMC Nurs. 2018 Apr 17;17:15. doi: 10.1186/s12912-018-0283-1. eCollection 2018.
- Pedersen T, Nicholson A, Hovhannisyan K, et al; Pulse oximetry for perioperative monitoring. Cochrane Database Syst Rev. 2014 Mar 17;(3):CD002013. doi: 10.1002/14651858.CD002013.pub3.
- Shiao SY, Ou CN; Validation of oxygen saturation monitoring in neonates. Am J Crit Care. 2007 Mar;16(2):168-78.
- Newborn pulse oximetry screening pilot update; Public Health England, 2019
- Jullien S; Newborn pulse oximetry screening for critical congenital heart defects. BMC Pediatr. 2021 Sep 8;21(Suppl 1):305. doi: 10.1186/s12887-021-02520-7.
- East CE, Begg L, Colditz PB, et al; Fetal pulse oximetry for fetal assessment in labour. Cochrane Database Syst Rev. 2014 Oct 7;(10):CD004075. doi: 10.1002/14651858.CD004075.pub4.
- Mitchell JM, Walsh S, O'Byrne LJ, et al; Association between intrapartum fetal pulse oximetry and adverse perinatal and long-term outcomes: a systematic review and meta-analysis protocol. HRB Open Res. 2024 Mar 28;6:63. doi: 10.12688/hrbopenres.13802.2. eCollection 2023.
- Asthma; NICE Klinische Wissenszusammenfassung. Januar 2025 (nur Zugang in Großbritannien)
- Richtlinien für das Management von ambulant erworbener Pneumonie bei Erwachsenen; Britische Thoraxgesellschaft (2009), Thorax Bd. 64 Sup III
- Lim WS, Smith DL, Wise MP, et al; British Thoracic Society community acquired pneumonia guideline and the NICE pneumonia guideline: how they fit together. BMJ Open Respir Res. 2015 May 13;2(1):e000091. doi: 10.1136/bmjresp-2015-000091. eCollection 2015.
- Martin D, Johns C, Sorrell L, et al; Effect of skin tone on the accuracy of the estimation of arterial oxygen saturation by pulse oximetry: a systematic review. Br J Anaesth. 2024 May;132(5):945-956. doi: 10.1016/j.bja.2024.01.023. Epub 2024 Feb 17.
- Jubran A; Pulse oximetry. Crit Care. 2015 Jul 16;19:272. doi: 10.1186/s13054-015-0984-8.
- Hanning CD, Alexander-Williams JM; Pulse oximetry: a practical review. BMJ. 1995 Aug 5;311(7001):367-70.
Lesen Sie unten weiter
About the authorView full bio

Dr Hayley Willacy, FRCGP
Allgemeinmediziner, Medizinischer Autor
MBChB (1992), DRCOG, DFFP, MRCOG (Part 1) MRCGP (2007), DFSRH (2013), MSc - medical education (2020)
Dr Hayley Willacy was an NHS GP working in northwest England, who retired from clinical practice in 2022 after 30 years.
About the reviewerView full bio

Dr Colin Tidy, MRCGP
Allgemeinmediziner, Medizinischer Autor
MBBS, MRCGP, MRCP (Paediatrics), DCH
Dr. Colin Tidy ist ein NHS-Arzt mit Sitz in Oxfordshire.
Artikelverlauf
Die Informationen auf dieser Seite wurden von qualifizierten Klinikern verfasst und begutachtet.
Nächste Überprüfung fällig: 16. Okt. 2027
17. Okt. 2024 | Neueste Version

Fragen, teilen, verbinden.
Durchsuchen Sie Diskussionen, stellen Sie Fragen und teilen Sie Erfahrungen zu Hunderten von Gesundheitsthemen.

Fühlen Sie sich unwohl?
Bewerten Sie Ihre Symptome online kostenlos